A High-Performance Thin Layer Chromatography Densitometric Method for the assay of Mebendazole Tablets

  • Vicky Manyanga 1Department of Medicinal Chemistry, Muhimbili University of Health and Allied Sciences, P.O. Box 65013, Dar Es Salaam, Tanzania
  • Emmanuel Kimaro 1Department of Medicinal Chemistry, Muhimbili University of Health and Allied Sciences, P.O. Box 65013, Dar Es Salaam, Tanzania
  • Joseph Sempompe 1Department of Medicinal Chemistry, Muhimbili University of Health and Allied Sciences, P.O. Box 65013, Dar Es Salaam, Tanzania
  • Eliangiringa Kaale 2Pharm R&D Laboratory, School of Pharmacy, Muhimbili University of Health and Allied Sciences P.O. Box 65013, Dar Es Salaam, Tanzania.
  • Mhina Chambuso Department of Medicinal Chemistry, Muhimbili University of Health and Allied Sciences, P.O. Box 65013, Dar Es Salaam, Tanzania.
Keywords: HPTLC, method validation, assay, mebendazole

Abstract

A simple, precise and accurate high-performance thin layer chromatographic (HPTLC) method was developed for the assay of mebendazole tablets. The separation was carried out by using HPTLC Silica gel 60 F254, (20 × 10 cm) with 250 μm thickness using ethyl acetate - ammonia 25% solution (25:0.5) as a mobile phase. HPTLC separation of the drug was followed by densitometry measurement at 310 nm. Mebendazole was satisfactorily resolved with retention factor (Rf) values of 0.51 ± 0.02. The method complied with International Conference on Harmonization acceptance criteria for linearity (250 - 600 ng), precision, accuracy and specificity.

Published
2020-09-03
How to Cite
Manyanga, V., Kimaro, E., Sempompe, J., Kaale, E., & Chambuso, M. (2020). A High-Performance Thin Layer Chromatography Densitometric Method for the assay of Mebendazole Tablets. The East and Central African Journal of Pharmaceutical Sciences, 23(2), 54-59. Retrieved from https://uonjournals.uonbi.ac.ke/ojs/index.php/ecajps/article/view/581